FAQ - JuLi Fluorescent Cell Analyzer


Questions

Answers

Q1
Is analysis software included?   A1 Yes, it is located on the included SD card.
Q2
Why is there only 1-channel for fluorescence?   A2

No, the JuLi is designed to eliminate moving parts which would be susceptible to corrosion over time in the hot and humid environment of  a cell culture chamber.  Additional LEDs for different channels would require a moving shuttle.

 

Q3
Is there an automatic focus?   A3

No.

 

Q4
Is the stage motorized for auto-indexing?   A4

No.

 

Q5
How long does the JuLi need to be in an incubator before I can start using it?   A5

We recommend that you turn on the JuLi 30-60 mintutes before use to allow the system to equillibrate. Otherwise images may start to black-out due to condensation on the mirror delivering the image. In addition, smaller incubators may see a short term rise in temperature when the JuLi is on as it does put it out heat so you may need time for your incubator to equillibrate

 

Q6
What is the size and resolution for the field of view?   A6

Normal Mode

Field of View : 1.7mm x 1.35mm

Resolution : 1280 x 1024 pixels

 

Zoom Mode

Field of View : 0.85mm x 0.675mm

Resolution : 640 x 512 pixels